Exfo OTDR 2 Manual de usuario Pagina 209

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 219
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 208
Description of Event Types
OTDR 201
Reflective Event (Possible Echo)
Reflective Event (Possible Echo)
This symbol indicates a reflective event that can be a real reflection or an
echo produced by another stronger reflection located closer to the source.
In the example above, the launched pulse hits the third connector, is
reflected back to the OTDR and reflected again into the fiber. It then
reaches the third connector a second time and is reflected once more
to the OTDR.
The application would therefore detect a reflective event located at
twice the distance of the third connector. Since this event is almost null
(no loss), and since its distance is a multiple of the third connector
distance, the application would interpret it as a possible echo.
A reflectance value is specified for reflective events (possible echo).
Reflected power
(dB)
Distance
(km)
OTDR connector
Second connector
Third connector
Reflective event (possible echo)
Vista de pagina 208
1 2 ... 204 205 206 207 208 209 210 211 212 213 214 ... 218 219