Referencia de instrucciones SCPI
OTDR 381
Product-Specific Commands—Description
:CONFigure[1..n]:ACQuisition:PULSe?
Description This query returns the current pulse setting.
*RST reverts this setting to default value.
Syntax :CONFigure[1..n]:ACQuisition:PULSe?
Parameter(s) None
Response Syntax <Pulse>
Response(s) Pulse:
The response data syntax for <Pulse> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the pulse, in seconds.
Example(s) CONF:ACQ 1310 NM,1250 M,10 NS
CONF:ACQ:PULS? Returns 1E–8
See Also CONFigure[1..n]:ACQuisition:WAVelength:LIST?
CONFigure[1..n]:ACQuisition:RANGe:LIST?
CONFigure[1..n]:ACQuisition:PULSe:LIST?