Exfo FTB-5600 Distributed PMD Analyzer Manual de usuario Pagina 128

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 146
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 127
Optimizing Measurements
122 FTB-5600
Use a Larger Number of SOPs and Wavelengths
Use a Larger Number of SOPs and Wavelengths
This will ensure that the instrument can access close to a maximum
available wavelength number from the instrument. More SOPs can also
improve the measurement accuracy for RMS DGD in the available
wavelength range.
A maximum wavelength range and as many SOPs as possible for the
acquisition will reduce the uncertainty of a measurement (that is, to
achieve a minimum measurement uncertainty).
Note: At least 100 or 200 SOPs should be used for obtaining a meaningful
cumulative PMD curve as a function of distance, and more SOPs. For
example, 500 to 1000, is recommended in order to make a better and
clearer picture of cumulative PMD.
Vista de pagina 127
1 2 ... 123 124 125 126 127 128 129 130 131 132 133 ... 145 146

Comentarios a estos manuales

Sin comentarios